Product
SB500DM Conventional Diagnostic X-ray Machine
SB500DM Conventional Diagnostic X-ray Machine
 

SB500DM Conventional Diagnostic X-ray Machine

Product Type: Conventional diagnostic x-ray
Place of origin: G
Model No: SB500DM
Price Terms: FOB Guangzhou
Payment Terms: T/T.L/C.Western Union.MoneyGram
Package: Wooden case
Minimum Order: 1 set/sets
Delivery Time: 15 days after receive payment
Brand Name: SUNNY

1 Microcomputer control x-ray radiography assembles: 500mA and 630mA optional.

2 Equipped with double diagnostic table and x-ray rubes, tube voltage is 125kV, tube currency is 500mA, single direct bridge full-wave commutation.

3 X-ray tube capacity protection, anode starting-up protection and radiograph time-limit protection device semi-conduct controlled, fluoroscopy computed time device, x-ray tube over-hot protecionare provided, ensure x-ray tube run safely long time, improve accuracy of kV mA S.

4 Bigger scope of rotation diagnostic table, movable table muli-item spot-film device, easy to operate, radiation –proof, ensure operate and the tested.

5 Radiography table can move lengthways and crossover, can loaded all kinds of radiography dark box, use conveniently and it is beautiful.

6 Applied in all kinds of hospital for radiography such as general body,beam filtration,and chest.

7 With the X-TV system and topography device interface, the function of compartment TV remote control can be upgraded

Department

Technology parameters

High voltage generator control table and high voltage transform assemble

Fluoroscopy

Tube currency

0.3-5mA continuous currency

Tube voltage

50-110KV continuous voltage

Radiography

Tube currency

50-500mA shift/50-630mA shift

Tube voltage

45-125KV continuous voltage

Exposure time

0.02-5s,23 shift

Power

Voltage

380V±10%

Frequency

50Hz±1Hz

Capacity

30KVA

Inner resistance

0.3Ω

X-ray assembly

Type of x-ray

XD52-30.50/125

Focus

1×1.2×2 double focal

Rated rotate speed of anode

2800r/min

Changed beam limited device

Diagnosis table YS09D

Automatic, adjusted field of view

Radiograph table YS11S

Manually adjusted field of view, luminosity100LX

C15 Diagnostic table

Rotate of table

-15°~0°~90°automatic

Moving table

Automatic 700mm distance moving lengthways

Distance of focal table

400mm

High of table

820mm

Spot film device

Scope moving

Up and down

600mm

Left and right

±100mm

Forward and backward

300mm

Way of locking

Electromagnetism locking

Radiography item

5″×7

?

?

8″×10

?

?

10″×12

?

?

?

11″×14

?

CDG

Multi function radiography table

Moving table

Lengthways

±600mm

Electromagnetism locking

Across

±120mm

Mobile beam filtration grid

Grid density

28L/cm

Grid ratio

81

Focusing distance

100cm

Scope of moving

600mm

Radiography item

5″×7,8″×10,10″×12,11″×14,12″×15,14″×14,14″×17

Distance from surface to film

75mm

Height of bed

700mm

Moving pillar

±180°,per 90°orientation, moving lengthways 1500mm

Bulb tube

Rotate the cross arm±180ascend $ descend 1100mm

Option :

LSJ pillar radiography frame

The scope of ascend and descend

1450mm

The minimum distance from focal spot

380mm

Radiography item

8"×10 10"×12 11"×1412"×15

14"×14 14"×17

Fixed beam filtration role

Grid density

40L/cm

Grid radio

10:1

Focal distance

180cm

 

BACK