Product
S8500 Neonatal Monitor
S8500 Neonatal Monitor
 

S8500 Neonatal Monitor

Product Type: patient monitor
Place of origin: Guangdong
Model No: S8500
Price Terms: FOB Guangzhou
Payment Terms: T/T.L/C.Western Union.MoneyGram
Package: carton
Minimum Order: 1 set/sets
Delivery Time: 15 days after payment
Brand Name: SUNNY



Test Result

SunTech Advantage NIBP

Averaged deviation of Systolic pressure

-2.56 mmHg

Standard deviation of Systolic pressure

5.71 mmHg

Averaged deviation of diastolic pressure

-4.51 mmHg

Standard deviation of diastolic pressure

5.57 mmHg

 

Test Methods: Participants were 30 newborn baby who were measured 180 groups data 

 for systolic (SBP) and diastolic (DBP) blood pressure, each NIBP measurement reactivity 

to relevant IBP measure form indwelling catheter in umbilical artery or femoral artery , 

 Averaged deviation should be less than ±5mmHg and the standard deviation should be less than 8mmHg.

This clinical test methods follow the standard procedure of AAMI SP10:2002/A1:2003

 

 

BACK